Contact an Account ManagerA HPLC method for determination of flecainide in human plasma using fluorescence detection
Summary
Type: HPLC
Application #: 104660
Detection: fluorescence, excitation 300nm, emission 370nm
Conditions: eluent: 6.66g KH2PO4/l H2O + 4.8g H3PO4, 28% acetonitrile
flow rate 1.8ml/min
Authors: Humfeld S
Source: Chromatographia 28 (1989) 303 - 306
Details
Legend: Fig. 1. Chromatogram from a patient's plasma sample
1.) internal standard
2.) flecainide.